Picture this: You're following a protocol that says "centrifuge at 3,000 xg for 10 minutes." You walk over to your centrifuge, open it up, and look at the dial. It only shows RPM. No xg anywhere in sight.
Sound familiar?
If you've ever stood in front of a centrifuge feeling confused, you're definitely not alone. The xg-to-RPM conversion trips up everyone at some point, from students to seasoned researchers. But here's the good news: once you understand what's actually happening, it becomes surprisingly simple.
Today, we're going to demystify this conversion, explain why it matters, and give you the tools to nail it every single time.
The Playground Physics: What Is xg Anyway?
Let's start with the basics, and I mean really basic.
Remember being a kid on a merry-go-round? The faster it spun, the harder you had to hold on to avoid flying off. That outward push you felt? That's centrifugal force in action.
A centrifuge works exactly the same way. It spins your samples around in a circle, and that spinning creates a force that pushes everything outward, including the cells, proteins, or particles you're trying to separate.
Now, scientists needed a way to measure this force. They could have just said "spin it really fast" or "spin it kind of fast," but that's not very precise, is it?
So instead, they compare the centrifugal force to something we all understand: gravity.
The "g" in xg Stands for Gravity
When you drop something, gravity pulls it down to Earth at a specific rate. Scientists call that 1 g (one times the force of gravity).
When we say a centrifuge is spinning at 3,000 xg, we mean it's creating a force that's 3,000 times stronger than regular gravity. That "x" just means "times," so 3,000 xg literally means "3,000 times gravity."
Think about it this way: If gravity normally pulls things down at 1 g, then 3,000 xg is like having 3,000 Earths stacked on top of each other, all pulling at once. That's some serious force, and it's why your samples separate so quickly in a centrifuge.
Pretty cool, right?
So Why Do Centrifuges Show RPM Instead of xg?
Here's where it gets a little tricky.
RPM stands for Revolutions Per Minute. It's simply how many times the centrifuge rotor spins around in one minute. If your centrifuge is set to 3,000 RPM, the rotor completes 3,000 full circles every 60 seconds.
RPM is easy for the centrifuge to measure. It just counts how fast the motor is spinning. But here's the thing: RPM doesn't actually tell you how much force is acting on your samples.
The Bucket Analogy
Imagine swinging a bucket of water around your head in a circle. Now imagine two scenarios:
Scenario 1: You're swinging a bucket on a short rope, maybe half a meter long.
Scenario 2: You're swinging a bucket on a long rope, two meters long.
Even if you're spinning both buckets at the exact same speed (same RPM), the bucket on the longer rope experiences much more force. Why? Because it's traveling a greater distance with each spin. It's moving faster through space, even though it's completing circles at the same rate.
The same principle applies to centrifuges. The force (xg) depends not just on how fast the rotor spins (RPM), but also on how far from the center your samples are sitting. That's the rotor radius.
The Magic Formula: Converting RPM to xg
Alright, now that we understand the "why," let's get to the "how."
The formula that connects xg, RPM, and rotor radius looks like this:
xg = 1.118 × 10⁻⁵ × r × RPM²
Where:
xg = the relative centrifugal force (what you want to know)
r = the radius of the rotor in centimeters (distance from center to sample)
RPM = revolutions per minute (what your centrifuge dial shows)
Don't let the formula intimidate you. That weird number at the beginning (1.118 × 10⁻⁵) is just a constant that makes the math work out correctly. You don't need to understand where it comes from. You just need to plug in your numbers.
Breaking It Down Step by Step
Let's say your protocol tells you to spin samples at 3,000 xg, and you're using a centrifuge with a rotor radius of 10 cm. What RPM should you set?
We need to rearrange the formula to solve for RPM:
RPM = √(xg / (1.118 × 10⁻⁵ × r))
Now let's plug in our numbers:
RPM = √(3,000 / (1.118 × 10⁻⁵ × 10))
RPM = √(3,000 / 0.0001118)
RPM = √(26,833,929)
RPM ≈ 5,180
So you'd set your centrifuge to approximately 5,180 RPM to achieve 3,000 xg with a 10 cm rotor.
Why Does This Actually Matter in the Lab?
Okay, so now you know how to do the conversion. But why does it matter? Can't you just eyeball it?
Absolutely not. Here's why precision matters.
Different Speeds for Different Jobs
The force you use determines what gets separated and what stays suspended. Using the wrong force can completely ruin your experiment.
Low Speed (300-500 xg): Gentle Cell Pelleting
This is the gentle setting. At 300-500 xg, you're creating just enough force to pull cells down to the bottom of the tube without damaging them.
What it's used for:
Pelleting mammalian cells (like cultured HeLa cells or primary cells)
Collecting larger cells like yeast
Separating red blood cells from plasma
Any time you want intact, living cells at the end
Why this speed? Mammalian cells are fragile. Spin them too hard and you'll rupture their membranes, killing them and releasing all their contents into your sample. At 300-500 xg, gravity does the work gently. It's like asking someone to sit down versus pushing them into a chair.
Medium Speed (2,000-5,000 xg): Bacteria and Debris Removal
Now we're ramping up the force. At 2,000-5,000 xg, you're creating enough force to pull down smaller, denser particles.
What it's used for:
Pelleting bacterial cells (E. coli, for example)
Removing cell debris after lysis
Collecting precipitated proteins
Plasmid prep work
Clarifying lysates before protein purification
Why this speed? Bacteria are much smaller and denser than mammalian cells. They need more force to pellet efficiently. At 2,000-5,000 xg, bacteria come down quickly, and you can also pull down cellular debris. Those are the broken bits of cell membranes, organelles, and proteins that would otherwise cloud your sample.
The Reproducibility Problem
Here's the real kicker: protocols are written in xg for a reason. If a researcher in Tokyo develops a protocol using 3,000 xg, and you try to reproduce it using "approximately 5,000 RPM" without knowing your rotor radius, you might be applying way too much or too little force.
Science depends on reproducibility. Using xg standardizes the force across different centrifuges, different labs, and different countries. It removes the guesswork.
The Rotor Radius: The Variable You Can't Ignore
We keep talking about rotor radius, so let's address it directly.
The rotor radius is the distance from the center of the centrifuge rotor to where your sample sits. Different centrifuges have different rotor sizes, and this dramatically affects the conversion.
Common Rotor Sizes in the Lab
Here's what you'll typically encounter:
Small microcentrifuges (~8 cm radius): These are those little benchtop units you use for spinning down PCR tubes or quick spins of microcentrifuge tubes. Compact and convenient.
Standard benchtop centrifuges (~15 cm radius): The workhorses of most labs. These handle everything from 1.5 mL tubes to 50 mL conical tubes.
Larger centrifuges (~20 cm radius): These are the big units, often floor-standing models or refrigerated centrifuges used for processing large sample volumes.
Why does this matter? Because the same RPM on different rotors produces wildly different forces.
Let's see this in action. Imagine setting your centrifuge to 4,000 RPM:
On an 8 cm rotor: xg = 1.118 × 10⁻⁵ × 8 × 4,000² = 1,431 xg
On a 15 cm rotor: xg = 1.118 × 10⁻⁵ × 15 × 4,000² = 2,683 xg
On a 20 cm rotor: xg = 1.118 × 10⁻⁵ × 20 × 4,000² = 3,578 xg
Same RPM, dramatically different forces. This is why you absolutely must know your rotor radius.
How to Find Your Rotor Radius
Most centrifuge rotors have the radius printed right on them, usually on a label or etched into the metal. Look for something like "r = 15 cm" or "radius: 150 mm."
If you can't find it, check the centrifuge manual or the manufacturer's website. You can also measure it yourself with a ruler: measure from the center of the rotor shaft to the middle of the tube slot.
Once you know it, write it down and tape it to your centrifuge. Future you will thank present you.
A Quick Car Analogy to Wrap It Up
Think about driving a car around a curve. Three things affect how much you feel pushed to the side:
How fast you're going (that's RPM)
How tight the curve is (that's rotor radius. A tighter curve means more force)
The force you experience (that's xg)
If you go around a gentle curve at 50 mph, you barely feel anything. But take a sharp hairpin turn at the same speed, and you're gripping the wheel for dear life. Same speed, different curve, totally different experience.
Centrifuges work the same way. Same RPM, different rotor radius, completely different force.
This is especially important when separating blood components for hematocrit measurements or pelleting parasites in culture : https://adwoabiotech.blogspot.com/2025/06/how-to-calculate-hematocrit-for-malaria.html